Produkte/Ärztebedarf/Patientenmonitore

C90A Transport Monitor
New generation of specialized transport monitor

C90A Patientenmonitor

C90A is a Specialized transport monitor built on best quality ABS Plastic with And Shock Design so it can be used in Ambulance, patient Transport on Stretcher and ICU. It's More convenient in operation room for Laproscopic surgeries if used with twin combination of Masimo Spo2 & Etco2 (Respironics).

It's a versatile design with First Aid kit bag, which make the patient transport more convenient with twin battery option. C90A has a built in battery of 4.4 AH and the first aid kit contains 2.6AH making C90A work for continous 8 hours.

Features

  • 4.3 inch Color Touch Screen TFT, 1.2kg
  • 2. With Dual Battery Option of 4.4 AH & backup support of 2.6 AH, Monitor can support up to 8 hours of continuous monitoring; making it suitable for transportation of patient in Ambulance
  • 3. Flexible configuration to meet different clinical needs
  • Real-time parameters measurement display
  • Three level Audio and Visual Alarm
  • Powerful storage capacity:
    48 Hours trends graphic review
    1000 sets NIBP data review
    100 alarm records review
  • Real-time parameters measurement display with trends table

 

C90A Patientenmonitor Tasche
  • Waterproof and Anti-shock desin 
  • Portable First Aid Bag for easy carrying of Patient Monitor and first aid kit
  • Multiple carrying options: Shoulder/Stretcher/Ambulance
C90A Patientenmonitor Datenaustausch 
  • Can Exchange Data with C90 (Mother Unit)
  • Can be used as a Module for C90 (ECG/SP02/NIBP/TEMP/RESP/C02)
C90A Patientenmonitor Display 
  • Big Font
C90A Patientenmonitor Bedienung 
  • Comen Inherited feature of 9 common short cut keys
  • Multiple interfaces
C90A Patientenmonitor Anzeige 
  • Two channel Waveform Display

Technical Specifications
Safety
IEC6060M approved. CE marking according to MDD93/42/EEC

Performance Specifications
Display: 4.3 inch Color TFT touch screen
Power Supply: 100-240 VAC. 50/60HZ
Battery Type: Lithium-ion
Voltage: 15VDC
Capacitance 4.400 mAh

Dimension: 85 mm(W) x 190mm (L) x 150 mm (H)
Weight 1.2kg

ECG
Lead Type:  3-lead. 5-lead. 12-lead selectable
ECG waveform:  2 channels
Gain selection:  x0.125; x0.25; xO 5; x1; x2; auto
Sweep speed:  12.5mm/s, 25mm/s, 50mm/s
Heart Rate range Adult: 15-300bpm; Neonate/ Pediatric 15-350bpm
Accuracy: ±1bpm or ±1%.whichever is greater
Alarm range: 15-350 bpm

Respiration
Method: Thoracic impedance
Operation modes: Auto/Manual
Measurement range: Adult 0-120 bpm; Neonate/Pediatric 0-150 bpm
Alarm: YES, auditable and visual alarm, alarm events recallable

NIBP
Method: Automatic Oscillometric
Operation modes: Manual/Automatic/Continuous
Measurement types: Systolic, Diastolic, Mean
Range of systolic pressure: 
Adult Mode 40 - 270 mmHg
Pediatric Mode 40 - 200 mmHg
Neonatal Mode 40 - 135 mmHg
Range of diastolic pressure: 
Adult Mode 10 - 210 mmHg
Pediatric Mode 10-l50mmHg
Neonatal Mode 10-100 mmHg
Range of mean pressure: 
Adult Mode 20 - 230mmHg
Pediatric Mode 20-165mmHg
Neonatal Mode 20 -110mmHg
Accuracy:
Max mean error 5 mmHg
Max standard deviation 8 mmHg
Resolution: 1mmHg

COMEN SpO2
Measurement range: 0-100%
Resolution: 1%
Accuracy:
±2% (70-100%, Adult/Pediatric, non-motion)
±3% (70-100%, Neonate, non-motion)
±3% (70-100%, Adult/Pediatric/Neonate, motion)
0-69% unspecified
Alarm range: 1-100%

Masimo SpO2 (Optional)
Measurement Accuracy:
±2% (70-100%, Adu/Ped, Non-motion)
±3% (70-100%, Neo, Non-motion)
±3% (70-100%, Motion)

Nellcor SpO2 (Optional)
Measurement Accuracy:
±2% (70-100%, MAX-A, MAX-AL, MAX-N, MAX-P, MAX-I and MAX-FAST sensors);
±3% (70-100%, D-ys, DS-100A, OXI-A/N and OXI-P/I sensors);
0-69% unspecified

Puls rate
Range: 0-254bpm
Resolution: 1bpm
Accuracy: ±3bpm

Temperature
Accuracy: ±0.1°C (Excluding the sensor)
Measure range: 0-50°C
Channel: Dual-channel, Provide T1; T2;
Resolution: 0.1°C

Respironics EtCO2 (Optional Sidestream/ Mainstream)
Range: 0-150 mmHg
Accuracy:
±2% 0-40 mmHg
±5% 41 -70 mmHg
±8% 71 - 100 mmHg
±10% 101 - 150 mmHg
AwRR Accuracy: ±1rpm
Convenient desgin for intubated and non-intubated applications
Possible to work at low sample flow rate: 50ml/min

Convenient operation / Data

Configuration

ECH/HR

SPO2/PR

NIBP

TEMP

RESP

EtCO2

C90A (SpO2)

X

V

X

X

X

X

C90A (SpO2+NIBP)

X

V

V

X

X

X

C90A (ECG+SpO2)

V

V

X

X

X

X

C90A (ECG+SpO2+NIBP)

V

V

V

X

X

X

C90A (SpO2+ETCO2)

X

V

X

X

X

V

C90A (ECG+SpO2+ETCO2)

V

V

X

X

X

V

C90A (ECG+SpO2+NIBP+TEMP+RESP)

V

V

V

V

V

X

 



 