img/produkte/Haematologie/hem-o-test_vet.jpg

  • 3-part differentation of WBC, 23 parameters + 3 histograms
  • Low sample volume: 10μl blood
  • Unique dilute program, independent HGB detection system
  • Automatic probe cleaning and sample dilution
  • Data storage of 30000 samples with histograms


Technical specification

Parameters WBC, Lymph, Mid, Gran, Lymph%, Mid%, Gran%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, WBC Histogram, RBC Histogram, PLT Histogram, P-LCR.
Principles
  • Electrical resistance for counting and SFT method for hemoglobin.
  • Specification
  • Aperture Diameter 70 μm
  • 10 uL sample volume
  • Throughput: 35 samples/hour
  • Automatic and manual calibration
  • Automatic dilution when power on/off
  • High voltage cleaning after each measuremnet
  • QC method: X-B, L-J, SD, CV
  • Carryover
  • WBC<0.5%, RBC<0.5%, HGB<0.5%, PLT<2%
  • Operating Environment
  • Temperature 15°C - 30°C
  • Humidity 10% - 85%
  • Printout
  • Thermal recorder, printer
  • Optional external printer
  • Power Requirement175-265V - 50/60HZ
    Dimension390 x 460 x 400 mm (W*D*H)
    Net Weight20 kg