Multi-Analyte Linearity Test SetPH5020 INTENDED USE: Multi-Analyte Linearity Test Sets are for in vitro diagnostic use in verifying reportable ranges and determining linearity in automated, semi-automated and manual chemistry systems. The analytes included are Total Protein (TP), Albumin (ALB), Uric Acid (UA), Creatinine (CREAT), Glucose (GLU), Blood Urea Nitrogen (BUN), Lactate (LAC), Phosphorous (P), Sodium (Na+), Potassium (K+), Calcium (Ca++), Chloride (Cl-), Lithium (Li+), Iron (FE), Magnesium (Mg++), Cholesterol (CHOL), Triglycerides (TRIG), and Carbon Dioxide (CO2). Multi-Analyte Linearity Test Sets are designed to be compatible with all popular chemistry analyzers, providing clinicians with 6 dilution levels, and 3 ampules of 1ml each per level. They are manufactured such that a linear relationship exists between all levels. SUMMARY: Multi-Analyte Linearity Test Sets are used to establish the relationship between theoretical and actual performance of specified analytes. This control set will assist in the documentation of linearity, calibration verification and verification of linear range required by many inspection agencies. The control solutions can also be used to troubleshoot problems with chemistry systems, reagents, and / or calibration anomalies. INGREDIENTS: Purified chemicals for albumin, calcium, cholesterol, chloride, carbon dioxide, creatinine, glucose, iron, lactate, lithium, magnesium, phosphorus, potassium, sodium, total protein, triglyceride, urea nitrogen, and uric acid are stabilized and preserved in a bovine albumin solution. STORAGE AND STABILITY: When stored and refrigerated at 2 to 8° C, Multi-Analyte Linearity Test Sets are stable until the expiration date printed on the ampule or vial. Opened ampules must be used within the same working day or else discarded. Opened vials must be tightly capped and immediately returned to refrigeration after each use. Dispose if gross contamination is visible. INSTRUCTIONS FOR USE: Multi-Analyte Linearity Test Sets are ready-to-use, and require no reconstitution. Depending upon the range and sensitivity of your instrument's test method, you will be able to run a minimum of 4 prediluted levels, and a maximum of 6 for a specific analyte. Materials contained herein should be treated in the same manner as patient samples. If additional dilutions or pre-treatment are required as part of your testing procedure, please consult the user manual of your instrument's manufacturer. For Ampules: Before actual use, hold ampule by the top and shake gently. Then with light tapping, restore all liquid to the bottom. Break open carefully to avoid cutting of fingers - using the complementary ampule snapper provided with this test set. With pipette, aspirate liquid from ampule and transfer to one or more sample cups. For Vials: Gently mix the contents of each vial before sampling to ensure homogeneity. With pipette, aspirate liquid from vial and transfer to one or more sample cups. Replace cap immediately and store at 2 to 8°C. Duplicate or triplicate runs are advised when performing calibration verification. CALCULATION OF RESULTS: Users of our Linearity Test Sets are strongly advised to calculate their results via BGT free data reduction service. The computational method and graphical analyses deployed in our reports are far more rigorous than the manual procedures outlined below. In addition, we can save clinicians considerable time by performing all of their calculations for them at no cost. Simply enter data into our customized MS Excel spreadsheets, and email them to the address provided. Spreadsheets can be obtained by emailing us at the following address: info@biogentechnologies.de If performing calculations manually, however, the following considerations will apply. After sampling each level in duplicate or triplicate, calculate a Mean Recovered Value for each, and record in the worksheet space provided. Theoretical Values for each level can be obtained by multiplying the Mean Recovered Value of Level 4 with the "Linearity Factors" provided below. *Note: Given that low concentrations of sodium and chloride cannot be read by most chemistry analyzers, these analytes are targeted independently of other constituents. For this reason, the linear relationship they share is also different from other constituents. Linearity Factors for Sodium (Na+) and Chloride (Cl-) Level 1 0.500 Level 2 0.667 Level 3 0.834 Level 4 1.000 Level 5 1.334 Level 6 1.667 Linearity Factors for All Other Analytes Level 1 0.250 Level 2 0.500 Level 3 0.750 Level 4 1.000 Level 5 1.750 Level 6 2.500 SAMPLE CALCULATION: If the Mean Recovered value for Level 4, Glucose = 282mg/dL, you can calculate Theoretical Values by multiplying 282 by the "Linearity Factor" associated with each level. Example: Calculations:
| Theoretical Value | Mean Recovered Value | Level 1 = 282 x 0.125 | 35.3 | 36.1 | Level 2 = 282 x 0.375 | 105.8 | 107.2 | Level 3 = 282 x 0.625 | 176.3 | 175.3 | Level 4 = 282 x 1.000 | 282.0 | 282.0 | Level 5 = 282 x 1.750 | 493.5 | 493.1 | Level 6 = 282 x 2.500 | 705.0 | 707.2 |
In order to assess the linearity of a specific test method, plot results on standard linear graph paper using "Theoretical" as X-axis and "Recovered" as Y-axis. EXPECTED VALUES: Each lot of product is manufactured in such a way that a linear relationship exists between all levels. Actual results obtained may vary depending upon instrumentation and methodology used, as well as assay temperature. Results may also depend upon the accuracy of the instrument and reagent calibration. The degree of acceptable non-linearity is an individual judgment based upon a test analyte's methodology, clinical significance and medical decision levels. Technicians are advised to consult the analytical limits defined by the Clinical Laboratory Improvement Act of 1988 (CLIA '88). These criteria specify the total error allowed for most analytes in question, and they can be referenced at the following web address: www.biogentechnologies.de Analyte | Range | Albumin (ALB) | 0.6 - 6.0 g/dL | Total Protein (TP) | 1.2 - 12 g/dL | Cholesterol (CHOL) | 50 - 500 mg/dL | Triglycerides (TRIG) | 67.5 - 675 mg/dL | Calcium (CA) | 2 - 20 mg/dL | Glucose (GLU) | 70 - 700mg/dL | Phosphorous (P) | 1.2 - 12 mg/dL | Urea Nitrogen (BUN) | 13 - 130 mg/dL | Uric Acid (UA) | 1.5 - 15 mg/dL | Magnesium (Mg++) | 0.6 - 6.0 mg/dL | Potassium (K+) | 1.4 - 14 mmol/dL | Iron (Fe) | 71 - 710 μg/dL | Creatinine (CREAT) | 2.5 - 25 mg/dL | Lactate (LAC) | 12 - 120 mg/dL | Lithium (Li+) | 0.7 - 7.0 mmol/dL | Carbon Dioxide (CO2) | 4 - 40 mmol/dL | *Chloride (Cl-) | 48 - 160 mmol/dL | *Sodium (Na++) | 72 - 240 mmol/dL |
* When computing Theoretical Values for Chloride and Sodium, please be sure to use special linearity factors provided in this insert. The linear relationship for these two analytes is different from other constituents. SUPPLEMENTAL PRODUCTS: A zero or near zero point during your calibration verification studies is strongly advised. BGT manufactures an Aqueous-Based Zero-Dilution matrix that is compatible with this Multi-Analyte Linearity Test Set: Aqueous-Based Zero / Dilution Matrix Cat. No.: PH5051 Configuration: 10mL Dropper Bottle Note: This product can also be used as a zero or diluting factor for BGT ISE / Metabolite Linearity Test Set. REORDERING INFORMATION: Multi-Analyte Linearity Test Set Cat. No.: PH5020B Configuration: 6 x 3 x 1mL (Ampules) For technical assistance or to place an order, please call: Tel: 02551/4090 Fax: 02551/1298 Email: info@biogentechnologies.de (Please allow 3-7 days for delivery)

 | Hersteller: |  Günter Keul GmbH Von-Langen-Weg 10 D-48565 Steinfurt Tel.: 02551/2097 Fax.: 02551/80883 Web: www.keul.de
|
|